Wide Range Surveymeter Type WR707 designed & manufactured by Nucleonix Systems serves as a dual range two detector surveymeter. A sensitive detector (external) covers lower range of 1µR/hr to 10mR/hr & second detector (internal) covers 1mR/hr to 1R/hr. At a time one detector only can be used.
State-of-art electronic circuitry designed around a microcontroller chip give stable and accurate readings, on a dot-matrix LCD display. |
Features
- Conforms to ANSI N42.17A performance specifications for Health Physics instrumentation (portable) &
compliant.
- Instrument is compact, elegant and lightweight, and is designed around Microcontroller based electronic circuitry.
- The Instrument operates on an in-built battery supply of 6V DC. (dry cells)
- The Instrument can with-stand temp. upto 55oC & Rel. humidity upto 90%.
- Measures doserates upto 5R/hr
- Measures count rate in both CPS & CPM
- Accuracy +/-15% (Excluding background )
|
 |
| Specifications |
|
Radiation detected |
Gamma |
|
Detector |
A miniature built in halogen quenched GM Tube LND 713 with external energy compensating filter or its equivalent for high range and an external probe with a halogen quenched GM tube type : LND 78017 for lower range |
|
Ranges |
Low Range : 1µR/hr to 10mR/hr or 10nSv to 100 µSv/h
High Range : 1mR/hr to 5R/hr or 10 µSv/h to 50mSv/h |
|
Accuracy |
within +/-15% of readings |
|
Response time |
Less than 2 seconds. |
|
Time Constant |
Choice of 3 time constants LOW, MED, HI |
|
Aural |
Built in piezo buzzer beeps for every detected pulse. |
|
Display |
Dot matrix LCD display for reading both dose rate/CPM. |
|
Controls |
START, STOP, PROG, STORE/RECALL, INC & DEC control buttons provided on the front face of instrument to allow the user to program and operate the instrument in different modes. |
|
Data Storage |
Storage of readings in built-in EEPROM. Stored Data can be recalled on to the display or transmitted to PC |
|
Serial Port / Data Communication |
Built-in serial port facilitates data down loading into PC. Data Communication Software with connecting cable can be provided at extra cost as an additional option. |
|
Power supply |
7.5volts DC, DURACELL ULTRA or BPL Excell, Size AA, MN1500 LR6 (5 X 1.5Volts), Alkaline cells. |
|
Type test compliance |
a) Conforms to ANSI N42.17A performance specifications for Health Physics instrumentation (portable)
b) compliant |
|
Dimensions |
106W x 196L x 85 Ht in mm (Approx.) |