Digital Contamination Monitor (Micro) Type CM 710E designed, manufactured and supplied by Nucleonix Systems is a state-of-art, microcontroller based unit which essentially serves as a low level alarming survey meter cum Beta contamination monitor.
It measures contamination in CPS or CPM mode and low level gamma dose rate in mR/hr / µSv/h.
It has lot of unique features which include user interface, through a set of six command buttons & the visual indication to the user is on a 16x2 LCD dotmatrix display module. This unit has preset alarm facility which can be set throughout the range of the instrument. Unit can store upto 1000 readings and stored data can be downloaded into PC using data communication software, through built-in RS232 serial port. |
Features
- Conforms to ANSI N42.17A performance specifications for Health Physics instrumentation (portable) &
compliant.
- Microcontroller based design.
- Compact, Elegant and light weight.
- External probe with cover for Beta / Gamma selection.
- Measures doserate from 0-200 mR/h in 5 linear ranges.
- Measures dose rate in mR/h / mSv/h & countrate in cps/cpm.
- Accuracy +/-10% with Cs-137.
- Unit works on 7.5V, dry cells (5x1.5V)
- RTC for time stamp of each reading.
|
 |
| Specifications |
|
Detector |
(1) End window GM detector type GM 125 or its equivalent.
(2) Pancake G.M.Detector type GM 138 or equivalent. |
|
Detector Probe |
This consists of GM detector fixed inside a housing with removable cover for Beta measurement. Probe is fixed into a side clamp on one side of the instrument. |
|
Radiation detected |
Beta & Gamma |
|
Measurement Unit |
mR/h / mSv/h cps / cpm |
|
Measuring ranges |
(a) Doserate : 0.0mR/h to 100mR/h (or) 0-1000 mSv/h (b) Countrate : 0-99999 cps or 0-99999 cpm. |
|
Preset Alarm Range |
0-100mR/h (or) 0-1000 mSv/h (or) 0-99999 cps (or) 0-99999 cpm. |
|
Time Constant (TC) |
User has a choice of three TC’s LO (2 sec), MED (4 sec) & HI (8 sec). |
|
Over range |
Senses over range & shows on display. |
|
Audio |
Built-in piezo buzzer provides beeps for every detected pulse and alarm o/p for preset alarms. |
|
Accuracy |
+/- 10% |
|
Display Indication |
16x2 backlit LCD display for showing doserate / countrate apart from textual screens during configuration. |
|
Data Storage |
Can store upto 1000 readings. Stored data can be recalled on to display or transmitted to PC through serial port under program control. |
|
RTC |
Real time stamp is recorded along with each reading. |
|
User Interface |
START, STOP, PROG, STORE, INC & DEC are the six control buttons provided on the instrument. The above buttons allow the user to program & operate the instruments in appropriate manner. |
|
Serial Port / Data Communication |
Built-in serial port facilitates data down loading into PC. Data Communication Software with connecting cable can be provided at extra cost as an additional option. |
|
Fault Diagnostics |
Low battery indication is provided on the display. |
|
Power |
7.5 Volts, DC, size AA, MN1500 LR6 (5 X 1.5 Volts), Alkaline cells. |
|
Type test compliance |
a) Conforms to ANSI N42.17A performance specifications for Health Physics instrumentation (portable)
b) compliant |
|
Dimensions |
106W x 196L x 100Ht in mm (Approx.) |