Home AboutUs Products services news&events Dealers Education contactus
STATE OF ART INSTRUMENTATION FROM AN INNOVATIVE COMPANY
WIDE RANGE AREA MONITOR (1 mR/hr to 100 R/hr) [TYPE: WR 725I] Print
Technical Data
Wide Range Area Monitor type WR725I, manufactured by NUCLEONIX SYSTEMS employs state-of-art micro-controller based design and is primarily meant to serve as a Gamma Zone Monitor to indicate dose rates and give alarm, visual and aural, once the dose rates exceed the preset level fixed by the user. Also relays will be activated on alarm.
This unit will be useful for monitoring Gamma dose rate levels in working areas of radioisotope laboratories, radiology departments, medical & industrial radiological installations apart from its usefulness in Atomic Power Stations and other active labs.
This unit indicates the dose rate digitally on a 4 x 7 segment LED display. Each of the annunciator windows for NORMAL and ACTIVE conditions has LED array. Once alarm triggers ACTIVE window starts blinking.
Unit can be programmed / configured using front panel keypad which can be deactivated after completion of programming. Configuring the unit namely setting preset level, setting reset mode - AUTO/MANUAL etc are achieved by this keypad.
Unit also performs self-diagnostics for EHT failure, pulse processing electronics failure & detectors failure on power up.
Alarm acknowledge and reset pins are provided on the circular I/O connector for remote acknowledge & reset.
Features

  • Performance specifications confirm to ANSI N42.17A
  • EMI/EMC compliance as per IEC 61000.
  • Structural design seismic qualified confirming to IEEE-344 STD.
  • Instrument enclosure and detector assembly are IP-54 compliant.
  • Radiation tolerant upto 104 R.
  • Firmware tested based on IEEE std : 1012.
  • Microcontroller based design has been employed
  • Dose rate range covered 1mR/hr to 99R/hr.
  • Auto ranging & auto TC selection in the range of 16 sec to 0.5 sec depending upon dose rate
  • Large size 4x7 segment LED indication for dose rate is provided
  • Designed using an Ionization chamer LND52110
  • Large size WINDOWS for NORMAL & ACTIVE alarm condition
  • 4 to 20 mA current loop / 0-5V analog is available for remote indication on a I/O connector (optional)
  • 16x2 LCD display dotmatrix display for visualization & other parameters.
Specifications

Radiation to be detected : X -ray & Gamma Radiation.

Range : 1 mR/hr to 100 R/hr in Two Ranges
10 µSv/hr to 1000mSv/hr in Two Ranges
(Range extendable to 1000R/h)

Detector : Pressurized Ion Chamber (LND52110) with external energy compensating filter

Energy Dependence : Within +/- 15% with respect to Cs-137 in the range of 200 keV to 1.3 MeV.
Accuracy : +/ - 10% with Co-60 above 10mR/hr

EHT : -400 V to -700 V DC adjustable (Typical -500V)

Display : Auto Ranging direct reading, 4 digit 7 segment LED display & 16x2 LCD display. 4x7 LED display is interfaced using multiplexed display driver. 4x7 segment display is used for display of dose-rate and hardware status indication & 16x2 LCD for visualization of preset alarm and other parameters

Display Resolution : 1mR/hr (10 mSv/hr)

Overload : Senses overload above 200% of fullscale and upto 10R/h & indicates on display “OL”

Over-range : Senses if the radiation field being measured has exceeded the measurement range of the instrument and upto 200% of the instrument and displays “OFl”

Recorder output : 4 to 20 mA, with 600 ohm load.

Recorder output stability
a) Non-linearity : Max = 0.025% of Span
b) Offset current (Io=4mA) : Max = 0.0005% of Span / °C
c) Span Error (Io=20mA) : Max = 0.005% of Span / °C

Time Constant : First reading on Power ON within 5 secs.
Normal (Slow) : Within 4 sec in normal mode and within 2 sec in abrupt detection

Calibration Accuracy : +/- 10% through out the range.

Instrument “ON” Indication : Large Area Green LED Lamp. This will indicate the Normal condition also.

Alarm range : 1 mR/hr to 99 R/Hr (10 mSv/Hr to 999 mSv/Hr)
Range Extendable to 999 R/hr
The alarm level setting will be carried out through front panel keypad / RS485 port using handheld configurator / PC with password protection. Front panel keypad is provided with DIP switch de-activation

Alarm Indication :
a) Red (LED) flashing large area window display
b) Loud audio tone (dual frequency tone)

Alarm annunciation scheme : As tabulated below

Instrument Controls :
a) Acknowledgement switch for muting audio
b) Reset switch for resetting the Alarm indication and alarm relay.
c) Power ON/OFF switch (This is inside the cabinet) with Power ON indication.

Instrument Fault indication :
a) EHT failure: Visual alarm with flashing red LED indication & “Eht” message on display
b) Detector failure: Visual alarm with flashing red LED & “d-FL” message on display.
c) Microprocessor / microcontroller failure: Visual alarm with flashing green lamp.
d) Fault indications shall be cleared automatically if normal status is resumed.

Detector Housing :
• The Pressurized Ion Chamber is located external to the Monitor.
• It is housed in a suitable, air-tight SS shell and is connected to the Electronic Unit through a suitable Cable assembly (Length could be extended upto a maximum of 50 mtrs)
• The instrument is provided with 20 mtr cable between detector and the monitor.
• The detector housing will qualify minimum industrial protection Class IP-54.
• A separate bracket for wall mounting of the detector housing will be provided.

Monitor Enclosure :
• Vapour-tight, rugged & elegant.
• The door is provided with lock and key arrangement
• The enclosure shall qualify minimum industrial protection Class IP-54.
• Decorative with visual aesthetics, prominent alarm display and good readability.

Mounting: Detector housing is mounted using clamps on top of the monitor. The monitor is wall mountable type. Brackets for the monitor & detector housing shall be supplied along with the equipment.

Remote /External Console :
• 4 - 20 mA linear proportional to full scale display output. Current output will be able to drive load of 600 ohms. (Output circuitry shall be able to drive 200 mtrs.of twisted pair of wires).
• Two sets of potential free contacts of Alarm relay (Change over). Contact rating 3 Amp at 250 VAC. The relay shall be energized on normal condition and de-energised under alarm condition.
• Remote alarm acknowledgement and reset signals for the field instruments (Normally open contact).

• Indication of instrument fault condition (detector, EHT and LV supplies), over range & overload conditions by up-scale 4-20 mA. (22.5 mA).
• All these signals shall be terminated on a 17 pin socket (Allied Connectors). The corresponding mating plug with 5 mtr cable shall be supplied with the monitor.
• RS-485 serial / Ethernet port. This shall be in parallel with D-type connectors.

Computer interface : The monitor shall have a RS-485 Serial/ Ethernet Communication port for interfacing with a remote IBM PC-compatible computer. The features supported by Ethernet port are given below.
• The PC and the monitor shall operate in a host-slave configuration and the software protocol will be MODBUS-RTU / MODUB-TCP.
• The PC as the host shall give commands and send queries. The monitor will carry out various functions in response to the queries.
• The firmware of the monitor shall be able to send the instrument data like instrument ID, instrument type, input range, display range, alarm settings, alarm status, current reading, diagnostic status of EHT/GM tube etc. to the Host PC on demand.
• The firmware shall be able to receive commands from Host PC and carry out the setting of different parameters like instrument ID, instrument type, input range, display range, alarm settings, Ack, Reset, instrument address etc.

Self Diagnostics : The monitor shall have built-in self diagnostics. On being powered it shall perform tests to ensure that all components and sub systems are functioning properly. It shall check for the Power supply, High Voltage Supply, Detector and pulse processing electronics.

Input Power : 230VAC +/-10%, 50Hz, single phase supply. Power ON/OFF switch shall be provided with a neon indicator. Spike suppressor and line filter are also provided.
Environment : The instrument shall be able to withstand temperature upto 50 deg C and relative humidity upto 90% in radiation areas.
The instrument enclosure and detector assembly will comply with IP-54. Electronic units will withstand cumulative radiation dose of 10000 Rad. (30 years of operation).
Mechanical Enclosure:
Size : 357H x 380W x 140D
Weight : 8.5kg approx

Instrument Trolley : (offered separately)
A suitable stand for fixing area monitor will be supplied optionally. This will be made of MS and will be provided with brackets for mounting the instrument. This stand will be designed to confirm to Seismic tests. One mains supply board with required sockets, indicators and switches shall be provided.

Plot No.162 A & B, Phase-II, IDA, Cherlapally, Hyderabad-500051 India Phone:+91-40-27263701, 32918055, 32914548 Fax : 040 - 27262146
Copyrights (c) 2008, Nucleonix Systems, All rights reserved.