Single Channel Analyser Type : SC530 manufactured by NUCLEONIX is designed primarily to be a constituent unit of Modular Gamma Ray Spectrometer Type : GR611M.
Single Channel Analyser SC530 is essentially used for pulse height analysis. It has THREE modes of operations namely INTEGRAL, NORMAL & WINDOW. In Gamma Ray Spectrometer the scintillation detector pulse output is proportional to the Gamma Energy and hence accurate measurement of pulse height by Single Channel Analyser will give the energy information. |
Features
- Accepts both positive unipolar or bipolar pulses as input
- Modes : INTEGRAL, NORMAL & WINDOW modes of operation
- Discriminator : LLD & ULD
- Output : Positive TTL
- Window width constancy, better than 30mV
|
 |
| Specifications |
|
Input |
Positive unipolar or bipolar (0-10V) linear range. |
|
Input Impedance |
5K Ohm |
|
Pulse Pair Resolution |
Better than 1 micro second |
|
Output |
4V positive pulse of 0.5 micro sec width |
|
Output Impedance |
Less than 100 ohms
|
|
Lower Level Discriminator |
Variable from 0 to 10V by means of ten turn helipot. |
|
External LLD Reference Input |
(Optional ) 0 to 10V (12V maximum) 1K ohms input impedance |
|
LLD Output |
4V positive pulse of 0.5 micro sec width available on the rear panel |
|
LLD Output Impedance |
Less than 100 ohms |
|
Upper Level Discriminator |
Variable from (0-10V) in NORMAL mode. Variable from (0-1V) in WINDOW mode.
It is ineffective in INTEGRAL mode. |
|
Non-linearity |
Less than 0.3% of full scale for both discriminators. |
|
Window width Constancy |
Better than +/- 30 mV variation over the Linear 0-10V range. |
|
Maximum Zero Offset |
Less than 0.5% of full scale for both the discriminators. |
|
Power requirements |
+24V 20mA +12V 180mA
-12V 60mA -24V 5mA |
|
Dimensions |
Standard single width module |
|
Module connector |
Amphenol Connector Type : 26-159-24P-H (24 Pin Type) or NIM Standard as per AEC specifications TID 20893 (Rev) Type : AMP 204186-5 |