Home AboutUs Products services news&events Dealers Education contactus
STATE OF ART INSTRUMENTATION FROM AN INNOVATIVE COMPANY
POWER SUPPLY UNIT - [ Type : PS 402 ] Print
Technical Data

Power Supply PS402 designed for mating with NIM Instrumentation Bin IB402, offers a complete rack mountable assembly or can be housed for table-top use. The D.C. output voltages furnished by PS402 are +12V and -12V at upto 3A each and +24V and -24V at upto 1.5A each. A 115V a.c. power output is also distributed to the Bin. Power Supply PS402 is overload and short circuit protected and has excellent load and line regulation with low ripple.

Features

  • Generates +/-12V d.c. at 3 amps and +/-24V d.c. at 1.5 amp
  • Regulation better than +/-0.05%
  • Noise & ripple less than 3 mV
  • Built-in overload & short circuit protections
  • Output rating 96 watts
Specifications
Input
115V/230V AC, single phase 50Hz with 10% variation in input voltage range
DC output
+12V DC at 3A; -12V DC at 3A
+24V DC at 1.5A; -24V DC at 1.5A
total output load not to exceed 96W at 30o centigrade
AC output
115V AC @0.5A has been wiredup and provided at each of the module connectors.
Regulation
Better than +/-0.05% over combined range of zero to full load and input voltage of 88% to 110% of rated input at constant ambient temperature after a 60 min warmup.
Stability
After warmup the voltage shall not vary by more than +/-0.2% of the nominal values at any time during any 24 hours period at constant ambient temperature over the combined range of no load to full load and input voltage +10% to -12% of nominal
Temperature Range
0 deg. to +50 deg. C without derating. Operation upto +60 deg. C is possible with a derating of 3% per deg. C for temeratures above +50 deg. C.
Temperature Coefficient
Less than 0.02% per deg. C upto +50 deg C.
Noise & Ripple (rms)
The combined output voltage noise and ripple shall not exceed 3mV
Voltage Adjustment
Output is adjustable over a nominal range of atleast +/-0.5% by means of a screw driver adjustment. Adjustment potentiometers are provided on the PCBs and are easily accessible when top cover is removed.
Circuit Protection
a. Input of the mains supply is protected by two fuses.
b. Overload and short circuit protection is achieved by electronic circuitry. The output shall recover automatically when the overload / short circuit is removed. Continuous short circuit will not damage the power supply.
Dimensions
430W x 168D x 153H approx. (in mm)
Plot No.162 A & B, Phase-II, IDA, Cherlapally, Hyderabad-500051 India Phone:+91-40-32918055, 32914548 Fax : 040 - 27262146