Specifications
OUTPUT VOLTAGE 0V to 1000V continuously variable by a ten turn helipot provided on front panel. (All specifications are valid from 10V to 1000V).
OUTPUT INDICATION Provided on a 3 1/2 digit LED, DPM.
OUTPUT CONNECTOR MHV Socket (by default)
UHF Socket (if specified)
OUTPUT CURRENT It can deliver upto a maximum of 1mA.
OUTPUT POLARITY The unit is offered with a choice of both POSITIVE or NEGATIVE POLARITY selectable by reversing a polarity connector. (Provided inside).
LINE & LOAD REGULATION Better than 0.05% of full scale.
RIPPLE Less than 20 mV peak to peak.
PROTECTIONS Indefinite protection against overload and short circuit . Under overload the unit doesn’t meet the specifications whereas on removal of short circuit it will recover by itself.
POWER REQUIREMENT Draws power from a standard BIN / NIM Power Supply or MINIBIN power supply. +24V at 250mA +12V at 50mA -12V at 50mA
LINEARITY Better than +/- 0.2%
TEMPERATURE STABILITY Better than 100 PPM.
MECHANICAL DIMENSIONS Standard 2 bit.
REAR MODULE CONNECTOR Amphenol connector Type:26-159-24P-H (24 pin type) or NIM standard connector as per AEC specifications TID 20893 (Rev) Type : AMP 2041865 optional.