PAN CAKE DETECTOR - [ Type : GM 138 ] Print
Technical Data
This is a pan cake detector sensitive to Beta radiation and with window closed can be used for gamma doserate measurements also. Has large surface window area of 16 sq.cm.
Application
Beta-gamma Contamination monitoring and also for gamma doserate measurements

This Detector is a part of Beta/gamma pan cake contamination probe which is used with Contamination Monitor, portable as well as with Bench Model. This detector is equivalent to LND7311.

General
Gas filling : Ne + Halogen
Cathode Material : 446 SS
Maximum Length : 3.0 Inches
Effective Depth : 0.5 Inches
Maximum Dia : 2.11 Inches
Effective Dia : 0.61 Inches
Connector : Grid Cap
Operating temperature range : -55 to +75oC
Window
Areal Density : 20 mg/cm2
Effective Dia : 1.75 Inches
Material : Mica
Gamma Sensitivity
60 CPS / mR/hr ( Co60 )
Operating Characteristics
Max. Starting Voltage : 750 V
Recommended Operating voltage : 900 V
Operating voltage range : 850-1000V
Plot No.162 A & B, Phase-II, IDA, Cherlapally, Hyderabad-500051 India Phone:+91-40-27263701, 32918055, 32914548 Fax : 040 - 27262146