GM Detector type GM136 is a thin wall halogen quenched GM detector suitable for high doserate measurements.
General
Gas filling : Ne + Halogen
Admixture
Cathode Material : 446 SS
Maximum Length : 0.97 Inches
Effective Length : 0.13 Inches
Maximum Dia : 0.25 Inches
Effective Dia : 0.19 Inches
Operating temperature
range : -40 to +75oC
Wall
Areal Density : 90 mg/cm2
Thickness : 0.005 Inches
Gamma Sensitivity
0.35 CPS / mR/hr ( Co60 )
Operating Characteristics
Max. Starting Voltage : 400 V
Recommended
Operating voltage : 500 V
Operating voltage range : 450-600V
Note : Used in some of the Gamma Area Monitors manufactured by NUCLEONIX.
Plot No.162 A & B, Phase-II, IDA, Cherlapally, Hyderabad-500051 India Phone:+91-40-27263701, 32918055, 32914548 Fax : 040 - 27262146