Home AboutUs Products services news&events Dealers Education contactus
STATE OF ART INSTRUMENTATION FROM AN INNOVATIVE COMPANY
HIGH GAMMA G.M. DETECTOR - [ Type : GM 136E ] Print
Technical Data
GM 136E is a halogen quenched GM detector with an internal energy independent filter. This detector has an energy response that is flat within +/-20% over a range from 80 kev to 1.25 mev.

Used in some of the Gamma Area Monitors manufactured by NUCLEONIX.

This detector is offered as a component in HG726. This detector is quivalent to LND7165.

General
Gas filling : Ne + Halogen Admixture
Cathode Material : 446 SS
Maximum Length : 1.1 Inches
Effective Length : 0.25 Inches
Maximum Dia : 0.53 Inches
Effective Dia : 0.19 Inches
Operating temperature range : -40 to +75oC
Wall
Areal Density : 90 mg/cm2
Thickness : 0.005 Inches
Filter
Material : Sn
Thickness : 0.15 Inches
Gamma Sensitivity
0.35 CPS / mR/hr ( Co60 )
Operating Characteristics
Max. Starting Voltage : 380 V
Recommended Operating voltage : 500 V
Operating voltage range : 450-600V
Plot No.162 A & B, Phase-II, IDA, Cherlapally, Hyderabad-500051 India Phone:+91-40-27263701, 32918055, 32914548 Fax : 040 - 27262146
Copyrights (c) 2008, Nucleonix Systems, All rights reserved.