Radiography Survey Meter [ TYPE : RM752IP]

Technical Data
manual  

This product is designed around a low power Micro-controller Chip. It is provided with a Graphic LCD  display for indicating the dose-rate, cumulative dose etc. It covers wide range from 0.1 mR/hr to 50 R/hr with auto ranging. This unit has a facility for storing the data into built-in flash memory along with time stamp. The stored data can be retrieved Via Blue-tooth to mobile. Also, Readings can be 

Features
Specifications
Detector  GM counter Type LND714 with energy compensated filter.
 Radiation detected   X & Gamma, from 60 Kev to 1.3 MeV.
 Measurement Units   R/h / Sv/h /hr /cps/ cpm.
 Dose Rate   0.1 mR/hr – 50 R/hr 0.1 µSv/hr – 500 mSv/hr Cumulative Dose up to 1000 R or 10 Sv
Count rate   0 – 99999 CPS / CPM
 Preset alarm range  5 – 50R/hr (or) 50 – 500 mSv/hr.
 Energy Response  Within +/- 20% of true response from 60KeV to 1.25 Mev.
 Over range   Unit shows over range above 10% of full-scale range.
 Range Changeover  Automatic.
 Calibration Accuracy  +/- 15%.
 Display   Graphic LCD display with 128 X 64 Pixel and 46.5W X 27.6 Ht.
 Response Time   < 5 sec.
 Audio  Built-in piezo buzzer provides beeps for every detected pulse and continuous  alarm above for preset alarms, level.
 Data Storage Can store up to 2000 readings with time stamp. Stored data can be recalled back on to display or transmitted to mobile through Blue-tooth connection.
 Serial port / data communication Built – in Via Blue-tooth in wireless mode.
 User Interface Buttons  A power ON /OFF, PRG, SRT / ENT, STP / EXT, ▲ (INC), ▼ (DEC) are the Six control buttons allow the user to program and operate the instrument in desired manner.
 RTC  Real time stamp recorded with each reading, in data storage mode.
 Fault Diagnostics  Low battery indication is indicated on the display.
 Enclosure  Flame retardant ABS plastic case and IP56 Rated.
 Size  85W x 35H x 145D in mm (approx.)
 Weight  < 300 gms.
 Additional Options offered include  i) Pelican carrying case